Item | Content | Technical parameter |
Power supply | Voltage | 380V±38V |
Frequency | 50Hz±1Hz |
Capacity | ≥65kVA |
internal resistor | ≤0.17Ω |
Digital X-ray high voltage system (import high-frequency generator ) | photography | Power Output | 50KW |
Inverter Frequency | 30kHz |
Tube Voltage | 40kv—150kv step regulation |
Tube Current | 10mA—630mA step regulation |
Exposure time | 1.0s—6300ms step regulation |
Control interface | Touched LCD |
Control hand | One-button operation handle |
fluoroscopy | Tube Voltage | 40kv—120kv ontinuous regulation |
Tube Current | 0.5mA—6mA continuous regulation |
Digital Controlled X-ray Tube | Tube Brand | Toshiba |
Tube Focus:Large Focus/Small Focus | 1.2mm /0.6mm |
Input Power | Large focus: 50kW small focus: 19kW |
thermal capacity | 300 KHU |
Rotary anode speed | 2800rpm |
Micro-computer control remote diagnostic table | Material of the tabletop | high strength,low absorb carbon fiber |
Rotation of the table | 90°~0°~-25° |
Transverse travel of table | ±110mm |
Longitudinal travel of the photography holder and film spot device | ≥720mm |
X-ray Focus—film gauge | 1100 mm—1500 mm |
Rotating foot plate | ±360° rotation |
Full film photography size | 8″×10″—14″×17″ |
Compressor force | 80N-130N |
Dividing method for spot film | Full film, half dividing, three parts dividing, four parts dividing |
Control method for diagnostic table | Remote,table control,frequency conversion soft starting and stopping |
beam limiter | Electric Multi-leaf |
fixed bucky device for table | Grid density: 103L/INCH, Grid ratio: 10:1,focusing distance: 120cm,fixed type: 15″×18″ |
Intensifier and CRT Monitor | Image Intensifier | TOSHIBA (9″) Image Intensifier |
high resolution CCD (digital cameras) | ultra-low-light,standard 400,000 pixel |
Monitor | 14 medical high definition monitor, horizontal central resolution ratio :1000 lines,fringe: 800 lines,video bandwidth: 12.5MHz,50bit,image number/second: 25 frame |
CCU System | High definition line by line output mode, 8 level noise reduction,store 8 image;LIH(freeze the last frame),the image can be turned vertically and horizontally,positive and negative image; OSD(monitor show) |
DR Detector | Toshiba portable flat panel detector | Material CSI, space resolution≥3.7lp/mm, active area 350(H)×430(V), Pixels matrix 2448(H)×2984(V), Pixels space 143 μm |
DR acquisition workstation | Computer system | Brand: DELL OPTIPLEX 7010 business use desktop Processor: Core i5-3570 RAM: 4GB DDR3 Hard disk: 500GB 7200r, SATA Graphics: Intel GMA HD 3000 Integrated Intel 82579LM kilomega Ethernet LAN 10/100/1000 CD-ROM: Suitable for MT 16X variable speed DVD+/-RW, with double level read-in function Network card: BroadcomNetXtreme10/100/1000 PCIe fast Ethernet card. Graphics: 1GB AMD RADEON HD 7570, FH, with DVI-VGA Adapter Interface: PCIE serial and parallel port, Full-HD, MT |
LCD colorful screen | Dell professional P1913S 19" monitor |
DROC workstation software | DROC and software dongle |